<div class="csl-entry">Miedler, M. (2015). <i>Evaluation of the log ratio method as an alternative for quantitative micro x-ray fluorescence scanning measurements</i> [Diploma Thesis, Technische Universität Wien]. reposiTUm. https://doi.org/10.34726/hss.2015.27502</div>
</div>
-
dc.identifier.uri
https://doi.org/10.34726/hss.2015.27502
-
dc.identifier.uri
http://hdl.handle.net/20.500.12708/5180
-
dc.description
Abweichender Titel laut Übersetzung der Verfasserin/des Verfassers
-
dc.description.abstract
Energy-dispersive XRF analysis provides a fast, non-destructive way to determine element distributions of samples. Once a calibration is done, the method provides an almost effortless way to achieve quantitative information. However, these calibrations are most often not linear and do not pass the origin thus they are only applicable in a very narrow range around the used standards. With the 2008 presented log-ratio calibration equation (LRCE) a promising method that aimed to get rid of this non-linearity has been introduced. The current work investigates how well the method achieves this goal by measuring a set of seven standards with a micro-focused XRF setup and a commercially available hand-held XRF device from Bruker. Parameters goodness-of-fit R2, slope alpha and y-intercept beta have been calculated and compared with both XRF devices as well as the Compton peak correction and the raw data set. Prior to the micro XRF measurements the setup had to be rebuild due to a relocation, thus requiring a check of the system performance. This was done by a long-term drift measurement, ensuring that the same results are achieved in every measurement. Apart from these calibrations, measurements with the hand-held device have been carried out in the historic japanese tower in Brussels. The drift measurements pointed out a slight drift for high count rate elements but the error could be rendered insigni cant when correction methods are applied. The calculations of the micro XRF setup reveals that the LRCE correction is favorable to the Compton peak correction. The performance is especially good for poorly correlated elements. For the hand-held device only elements with low signal and strong air absorption perform better with the LRCE. The measurements of the historic site in Brussels could identify some of the applied materials an techniques on paper and wood basis.
en
dc.language
English
-
dc.language.iso
en
-
dc.rights.uri
http://rightsstatements.org/vocab/InC/1.0/
-
dc.subject
x-ray fluorescence
en
dc.subject
calibration
en
dc.subject
geology
en
dc.title
Evaluation of the log ratio method as an alternative for quantitative micro x-ray fluorescence scanning measurements
en
dc.title.alternative
Evaluierung der Log-Ratio Methode als Alternative zu quantitativen Rastermessungen mittels Röntgenfluoreszenzanalyse
de
dc.type
Thesis
en
dc.type
Hochschulschrift
de
dc.rights.license
In Copyright
en
dc.rights.license
Urheberrechtsschutz
de
dc.identifier.doi
10.34726/hss.2015.27502
-
dc.contributor.affiliation
TU Wien, Österreich
-
dc.rights.holder
Martin Miedler
-
tuw.version
vor
-
tuw.thesisinformation
Technische Universität Wien
-
tuw.publication.orgunit
E164 - Institut für Chemische Technologien und Analytik
-
dc.type.qualificationlevel
Diploma
-
dc.identifier.libraryid
AC12225560
-
dc.description.numberOfPages
117
-
dc.identifier.urn
urn:nbn:at:at-ubtuw:1-79303
-
dc.thesistype
Diplomarbeit
de
dc.thesistype
Diploma Thesis
en
dc.rights.identifier
In Copyright
en
dc.rights.identifier
Urheberrechtsschutz
de
tuw.advisor.staffStatus
staff
-
item.fulltext
with Fulltext
-
item.cerifentitytype
Publications
-
item.mimetype
application/pdf
-
item.openairecristype
http://purl.org/coar/resource_type/c_bdcc
-
item.languageiso639-1
en
-
item.openaccessfulltext
Open Access
-
item.openairetype
master thesis
-
item.grantfulltext
open
-
crisitem.author.dept
E164 - Institut für Chemische Technologien und Analytik